MD 640:

  • v012.015.4.003.065 release date 08/24/2017
    new: method 113 (Chlorine MR PP):
    determination of free, total and combined Chlorine very similar to method 110 but up to 3.5 mg/l Cl2
    by using DPD powder packs order no. 530180 (free) and 530190 (total)
    range: 0.02 – 3.5 mg/l Cl2
    method 376 (Anionic Surfactants)
    determination of Anionic Surfactants, adapted from Merck Spectroquant(R) No. 1.02552.0001, Cell Test
    range: 0.05 – 2 mg/l SDSA (MBAS calculated as sodium 1-dodecanesulfonate)
    0.06 – 2.56 mg/l SDBS ((MBAS calculated as sodium dodecylbenzenesulfonate)
    0.05 – 2.12 mg/l SDS ((MBAS calculated as sodium dodecyl sulfate)
    0.08 – 3.26 mg/l SDOSSA ((MBAS calculated as Sodium dioctyl sulfosuccinate)
    method 377 (Nonionic Surfactants)
    determination of Nonionic Surfactants, adapted from Merck Spectroquant(R) No. 1.01787.0001, Cell Test
    range: 0.10 – 7.5 mg/l Triton (R)X-100
    range: 0.11 – 8.25 mg/l NP 10 (Nonylphenol 10)
    method 378 (Cationic Surfactants)
    determination of Cationic Surfactants, adapted from Merck Spectroquant(R) No. 1.01764.0001, Cell Test
    range: 0.05 – 1.5 mg/l CTAB
    method 380 TOC LR (Total Organic Carbon)
    determination of Total Organic Carbon, adapted from Merck Spectroquant(R) No. 1.14878.0001, Cell Test
    range: 5.0 – 80.0 mg/l TOC
    method 381 TOC HR (Total Organic Carbon)
    determination of Total Organic Carbon, adapted from Merck Spectroquant(R) No. 1.14879.0001, Cell Test
    range: 50 – 800 mg/l TOC
    method 501 PTSA 2P (PTSA Fluorescence)
    PTSA Fluorescence measurement method including two point calibration
    range: 10 – 400 ppb PTSA
    method 511 Fluorescine 2P (PTSA Fluorescence)
    PTSA Fluorescence measurement method including two point calibration
    range: 10 – 300 ppm Fluorescine
    improved: enhanced precision at high concentrations by using method 400 (Zinc T).
    COD methods 130, 131 and 132 show now underrange if a result is below the detection limit of the test.
    Ranges are now:
    COD LR: 3 – 150 mg/l COD
    COD MR: 20 – 1500 mg/l COD
    COD HR: 0,2 – 15 g/l COD
MD 610:

  • v012.013.4.003.065 release date 08/24/2017
    new: method 113 (Chlorine MR PP):
    determination of free, total and combined Chlorine very similar to method 110 but up to 3.5 mg/l Cl2
    by using DPD powder packs order no. 530180 (free) and 530190 (total)
    range: 0.02 – 3.5 mg/l Cl2
    method 376 (Anionic Surfactants)
    determination of Anionic Surfactants, adapted from Merck Spectroquant(R) No. 1.02552.0001, Cell Test
    range: 0.05 – 2 mg/l SDSA (MBAS calculated as sodium 1-dodecanesulfonate)
    0.06 ñ 2.56 mg/l SDBS ((MBAS calculated as sodium dodecylbenzenesulfonate)
    0.05 ñ 2.12 mg/l SDS ((MBAS calculated as sodium dodecyl sulfate)
    0.08 ñ 3.26 mg/l SDOSSA ((MBAS calculated as Sodium dioctyl sulfosuccinate)
    method 377 (Nonionic Surfactants)
    determination of Nonionic Surfactants, adapted from Merck Spectroquant(R) No. 1.01787.0001, Cell Test
    range: 0.10 – 7.5 mg/l Triton (R)X-100
    range: 0.11 – 8.25 mg/l NP 10 (Nonylphenol 10)
    method 378 (Cationic Surfactants)
    determination of Cationic Surfactants, adapted from Merck Spectroquant(R) No. 1.01764.0001, Cell Test
    range: 0.05 – 1.5 mg/l CTAB
    method 380 TOC LR (Total Organic Carbon)
    determination of Total Organic Carbon, adapted from Merck Spectroquant(R) No. 1.14878.0001, Cell Test
    range: 5.0 – 80.0 mg/l TOC
    method 381 TOC HR (Total Organic Carbon)
    determination of Total Organic Carbon, adapted from Merck Spectroquant(R) No. 1.14879.0001, Cell Test
    range: 50 – 800 mg/l TOC
    improved: enhanced precision at high concentrations by using method 400 (Zinc T).
    COD methods 130, 131 and 132 show now underrange if a result is below the detection limit of the test.
    Ranges are now:
    COD LR: 3 – 150 mg/l COD
    COD MR: 20 – 1500 mg/l COD

    COD HR: 0,2 – 15 g/l COD

  • v012.013.4.003.058
    release date 08/17/2015
    new: method 224 (fe in mo):
    determination of all dissolved iron and unsolved iron in the presence of high molybdate concentrations
    range: 0.01 – 1.80 mg/l Fe
  • v012.013.4.003.057 release date 05/21/2015
    improved: enhanced precision at very low concentrations by using methods 110, 122 and 81
    removed: method 63 (monochloramine), because chemicals are no longer available
MD 600/ MaxiDirect:

  • v012.007.3.003.065 release date 08/24/2017
    new: method 113 (Chlorine MR PP):
    determination of free, total and combined Chlorine very similar to method 110 but up to 3.5 mg/l Cl2
    by using DPD powder packs order no. 530180 (free) and 530190 (total)
    range: 0.02 – 3.5 mg/l Cl2
    method 376 (Anionic Surfactants)
    determination of Anionic Surfactants, adapted from Merck Spectroquant(R) No. 1.02552.0001, Cell Test
    range: 0.05 – 2 mg/l SDSA (MBAS calculated as sodium 1-dodecanesulfonate)
    0.06 – 2.56 mg/l SDBS ((MBAS calculated as sodium dodecylbenzenesulfonate)
    0.05 – 2.12 mg/l SDS ((MBAS calculated as sodium dodecyl sulfate)
    0.08 – 3.26 mg/l SDOSSA ((MBAS calculated as Sodium dioctyl sulfosuccinate)
    method 377 (Nonionic Surfactants)
    determination of Nonionic Surfactants, adapted from Merck Spectroquant(R) No. 1.01787.0001, Cell Test
    range: 0.10 – 7.5 mg/l Triton (R)X-100
    range: 0.11 – 8.25 mg/l NP 10 (Nonylphenol 10)
    method 378 (Cationic Surfactants)
    determination of Cationic Surfactants, adapted from Merck Spectroquant(R) No. 1.01764.0001, Cell Test
    range: 0.05 – 1.5 mg/l CTAB
    method 380 TOC LR (Total Organic Carbon)
    determination of Total Organic Carbon, adapted from Merck Spectroquant(R) No. 1.14878.0001, Cell Test
    range: 5.0 – 80.0 mg/l TOC
    method 381 TOC HR (Total Organic Carbon)
    determination of Total Organic Carbon, adapted from Merck Spectroquant(R) No. 1.14879.0001, Cell Test
    range: 50 – 800 mg/l TOC
    improved: enhanced precision at high concentrations by using method 400 (Zinc T).
    COD methods 130, 131 and 132 show now underrange if a result is below the detection limit of the test.
    Ranges are now:
    COD LR: 3 – 150 mg/l COD
    COD MR: 20 – 1500 mg/l COD
    COD HR: 0,2 – 15 g/l COD
  • v012.007.3.003.058
    release date 08/17/2015
    new: method 224 (fe in mo):
    determination of all dissolved iron and unsolved iron in the presence of high molybdate concentrations
    range: 0.01 – 1.80 mg/l Fe
  • v012.007.3.003.057
    release date 05/21/2015
    improved: enhanced precision at very low concentrations by using methods 110, 122 and 81
    removed: method 63 (monochloramine), because chemicals are no longer available
MultiDirect:

  • v012.012.2.003.065 release date 08/24/2017
    new: method 113 (Chlorine MR PP):
    determination of free, total and combined Chlorine very similar to method 110 but up to 3.5 mg/l Cl2
    by using DPD powder packs order no. 530180 (free) and 530190 (total)
    range: 0.02 – 3.5 mg/l Cl2
    method 376 (Anionic Surfactants)
    determination of Anionic Surfactants, adapted from Merck Spectroquant(R) No. 1.02552.0001, Cell Test
    range: 0.05 – 2 mg/l SDSA (MBAS calculated as sodium 1-dodecanesulfonate)
    0.06 – 2.56 mg/l SDBS ((MBAS calculated as sodium dodecylbenzenesulfonate)
    0.05 – 2.12 mg/l SDS ((MBAS calculated as sodium dodecyl sulfate)
    0.08 – 3.26 mg/l SDOSSA ((MBAS calculated as Sodium dioctyl sulfosuccinate)
    method 377 (Nonionic Surfactants)
    determination of Nonionic Surfactants, adapted from Merck Spectroquant(R) No. 1.01787.0001, Cell Test
    range: 0.10 – 7.5 mg/l Triton (R)X-100
    range: 0.11 – 8.25 mg/l NP 10 (Nonylphenol 10)
    method 378 (Cationic Surfactants)
    determination of Cationic Surfactants, adapted from Merck Spectroquant(R) No. 1.01764.0001, Cell Test
    range: 0.05 – 1.5 mg/l CTAB
    method 380 TOC LR (Total Organic Carbon)
    determination of Total Organic Carbon, adapted from Merck Spectroquant(R) No. 1.14878.0001, Cell Test
    range: 5.0 – 80.0 mg/l TOC
    method 381 TOC HR (Total Organic Carbon)
    determination of Total Organic Carbon, adapted from Merck Spectroquant(R) No. 1.14879.0001, Cell Test
    range: 50 – 800 mg/l TOC
    improved: enhanced precision at high concentrations by using method 400 (Zinc T).
    COD methods 130, 131 and 132 show now underrange if a result is below the detection limit of the test.
    Ranges are now:
    COD LR: 3 – 150 mg/l COD
    COD MR: 20 – 1500 mg/l COD
    COD HR: 0,2 – 15 g/l COD
  • v012.002.3.003.058
    release date 08/17/2015
    new: method 224 (fe in mo):
    determination of all dissolved iron and unsolved iron in the presence of high molybdate concentrations
    range: 0.01 – 1.80 mg/l Fe
  • v012.002.3.003.057
    release date 05/21/2015
    improved: enhanced precision at very low concentrations by using methods 110, 122 and 81
    removed: method 63 (monochloramine), because chemicals are no longer available
PM 620/ PoolDirect:

  • v012.012.12.003.065 release date 08/24/2017
    new: method 113 (Chlorine MR Powder):
    determination of free, total and combined Chlorine very similar to method 110 but up to 3.5 mg/l Cl2
    by using DPD powder packs order no. 530180 (free) and 530190 (total)
    range: 0.02 – 3.5 mg/l Cl2
  • v012.012.3.003.057
    release date 05/21/2015
    improved: enhanced precision at very low concentrations by using method 110
PM 630:

  • v012.012.14.003.065 release date 08/24/2017
    new: method 113 (Chlorine MR Powder):
    determination of free, total and combined Chlorine very similar to method 110 but up to 3.5 mg/l Cl2
    by using DPD powder packs order no. 530180 (free) and 530190 (total)
    range: 0.02 – 3.5 mg/l Cl2
    v012.012.14.003.057 release date 05/21/2015
    improved: enhanced precision at very low concentrations by using method 110 (Chlorine Powder)
SpectroDirect
  • v201.001.1.001.023 release date 08/24/2017
    new: method 113 (Chlorine MR PP):
    determination of free, total and combined Chlorine very similar to method 110 but up to 3.5 mg/l Cl2
    by using DPD powder packs order no. 530180 (free) and 530190 (total)
    range: 0.02 – 3.5 mg/l Cl2method 376 (Anionic Surfactants)
    determination of Anionic Surfactants, adapted from Merck Spectroquant(R) No. 1.02552.0001, Cell Test
    range: 0.05 – 2 mg/l SDSA (MBAS calculated as sodium 1-dodecanesulfonate)
    0.06 – 2.56 mg/l SDBS ((MBAS calculated as sodium dodecylbenzenesulfonate)
    0.05 – 2.12 mg/l SDS ((MBAS calculated as sodium dodecyl sulfate)
    0.08- 3.26 mg/l SDOSSA ((MBAS calculated as Sodium dioctyl sulfosuccinate)

    method 377 (Nonionic Surfactants)
    determination of Nonionic Surfactants, adapted from Merck Spectroquant(R) No. 1.01787.0001, Cell Test
    range: 0.10 – 7.5 mg/l Triton (R)X-100
    range: 0.11 – 8.25 mg/l NP 10 (Nonylphenol 10)

    method 378 (Cationic Surfactants)
    determination of Cationic Surfactants, adapted from Merck Spectroquant(R) No. 1.01764.0001, Cell Test
    range: 0.05 – 1.5 mg/l CTAB

    method 380 TOC LR (Total Organic Carbon)
    determination of Total Organic Carbon, adapted from Merck Spectroquant(R) No. 1.14878.0001, Cell Test
    range: 5.0 – 80.0 mg/l TOC

    method 381 TOC HR (Total Organic Carbon)
    determination of Total Organic Carbon, adapted from Merck Spectroquant(R) No. 1.14879.0001, Cell Test
    range: 50 – 800 mg/l TOC
    improved: enhanced precision at high concentrations by using method 400 (Zinc T).

    COD methods 130, 131 and 132 show now underrange if a result is below the detection limit of the test.
    Ranges are now:
    COD LR: 3 – 150 mg/l COD
    COD MR: 20 – 1500 mg/l COD
    COD HR: 0,2 – 15 g/l COD